Wastewater UV% Transmission Analyser – Accuview

Description:

This fourth generation Wastwater UV% Tra

SKU: HF19571 Categories: , ,

Product Description

Description:

This fourth generation Wastewater UV% Transmission Analyzer has been developed specifically for the harsh environment of wastewater applications. The AccuView Wastewater is manufactured using advanced technology, a NEMA 4X stainless steel enclosure and revolutionary ultrasonic technology making it the most reliable, maintenance free and cost efficient UV5 analyser for wastewater. The Auto Clean Ultrasonic cleaning system continuously cleans the optical chamber providing a maintenance free reliable environment.

Applications:

Waste (effluent) water in industry or municipalities.

Features:

  • Optical design – New optical design allows consistent readings with laboratory %T analysers
  • Bellows Pump – Pump included to provide reliable sample flow to the instrument
  • Certifiable traceable standards – HF Scientific certifies the 100%T calibration standard is prepared as outlined in the Standard Methods for the Examination of Water and Wastewater, 21st edition, Method 1080 A-C, Methods for Preparation of Reagent Water
  • Backlight Display – user adjustable backlit display allows viewing in low light conditions. Stainless steel enclosure includes large viewing window to allow monitoring of readings without opening the enclosure
  • Certification – Listed or certified to CE, UL, CSA (ETL, ETLc)

Guarantee:

1 year  
Product Specification
Description
Range
0 – 100% transmission (Auto ranging)
Resolution
0.1% T
Repeatability
approx. 0.1% T
Accuracy
approx. 0.1% T
Wavelength
Ultraviolet 253.7 nm
Response Time
User selectable updates from 10 – 60 minutes
Cleaning System
Ultrasonic
Operating Temperature
-20 to 50 degrees C – low temperature operation requires optional cold weather kit
Standard Outputs
4 – 20 mA with isolator or optional RS485 with modbus protocol
Alarm
user settable 240 VAC form C relay
Sample Temperature
0 – 40 degrees C
Storage Temperature
-20 to 60 degrees C
Pressure / Flow rate
Mximum 200 psi / 1000 ml per minuteBuilt in regulation provided
Sample Lift
3.5 m
Sample Strainers
Foot Strainer and T-Strainer
Pump Type
Bellows
Flow Rate
approx. 500 ml / minute
Display
multiline custom LCD with backlight
Power Requirements
100 to 240 VAC 47 – 63 Hz
Enclosure
stainless steel with lexan window
NEMA ratings
Desinged to meet NEMA 4X
Agency testing
CE, CE safety self declare, ET Lus and ET Lc
Dimensions
50.8 x 50.8 x 33.2 cm
Shipping weight
22.6 kg

Content:

  • AccuView OnLine UV% instrument 100 – 240 VAC, for pressurised or gravity feed systems
  • 4 – 20 mA output with isolator
  • 2 alarm contacts
  • inline flow and pressure regulator
  • dessicant
  • power supply
  • 100% T standard
  • spare cuvette with ultrasonic transducer
  • operator’s manual

 

Additional Information

Weight 22.60 kg