Turbidimeter – PVT Process

Product Description

NEW INNOVATION! The PVT Process Turbidimeter from Lovibond® Tintometer® has an advanced design and communication system that has been user designed to make life easier for water process controllers.

One of the most vital parameters for healthy drinking water is turbidity. Why is turbidity important?
Today, an operator is able to view the turbidity of process water in real time, with new technology that is especially designed to be user-friendly – the Lovibond® PVT Process Turbidimeter and a smart device.

Years of development by Lovibond® Tintometer® have resulted in this advanced turbidimeter. A team of globally recognised turbidity experts was gathered to find out what issues they experienced with current turbidity systems. A user inspired system was then developed.

Lovibond® Tintometer® created a secure system that greatly reduces complexity by allowing a user to interact with an unlimited number of turbidimeters using a single mobile device app. There is no need for a dedicated controller for each instrument.

A maximum of three clicks will take you anywhere you need to be – making it much easier to ensure that turbidity values are compliant with regulations.

pvt100-turbidimeter

Applications:

  • Process water
  • Drinking water

Features:

  • Two ways to interact with the sensor: use the touch screen, or connect to the sensors using the AquaLXP® App (iOS and Android)
  • Intuitive screen – large, easy-to-read touch screen prompts for routine operations and procedures, such as calibration
  • AquaLXP® App can calculate statistics, integrate operator notes, access instructions and maintenance logs
  • Records: a record of every reading, calibration, verification, cleaning or other maintenance performed over the lifetime of the sensor is automatically logged into the instrument history
  • Security: a certain physical proximity is required, in addition to two layers of access passwords
  • Alarms: communicated on instrument, mobile device and SCADA system
  • Accurate and ultra-stable measurements: unsurpassed low range accuracy (below 1 NTU) and patent-pending bubble exclusion system
  • Heated optical assembly – eliminates the need for desiccants with no chance for condensation and fogging
  • Easy clean flow body – no nooks or crannies, "quick-connect" features for easy draining during calibrations and cleaning, no need to disconnect tubing to access a sample
  • Low volume flow body – quick response rate to turbidity spikes and uses much less water and calibration standards. 70% less volume than competitive units. Optimal flow rate of 50 – 80 ml / minute
  • Light: choose from infra-red or white LED light source
  • T-CALplus™ Turbidity Standards: long shelf life, no degradation in cold or freezing environments. T-CALplus™ packaging is vacuum sealed for easy mixing without making bubbles. Also available in 500 ml bottles in different concentrations for verification and calibration
  • Compliance: ISO 7027; EPA Approval pending

Guarantee:

One year.

Measurement range 0.0001 to 100 NTU
Accuracy ± 2% of reading from 0 to 10 NTU
± 4% of reading between 10 and 100 NTU
Stray light PTV 1000 IR (ISO): < 0.005 / 5 mNTU
PTV 1000 WL (EPA*): < 0.015 / 15 mNTU
PTV 2000 RL (EPA*): < 0.008 / 8 mNTU
Limit of detection PTV 1000: < 0.0005 NTU
PTV 2000: < 0.0001 NTU
Limit of quantitation PTV 1000: better than 0.005 NTU
PTV 2000: better than 0.001 NTU
Displayed resolution up to 0.0001 NTU (range dependent)
or 5 digits displayed
Repeatability / Precision Better than 1% at 1 NTU
Initial response 10% change: 15 seconds @ max. flow
Step response T-90
Signal averaging User selectable: 1, 3, 6, 10, 30, 60 and 90 seconds defaulted to 30 seconds
Sample temperature 0 to 50°C
Max. sample temperature: 70°C
Sample flow 30 to 500 ml / minute
Optimal flow: 50 – 80 ml / minute
Operating pressure Atmosphere
Ambient / Operating temperature range 5 to 50°C
Ambient / Operating humidity range 5 to 95%
Storage and shipping temperature -40 to 60°C
Power requirements 90 to 264 VAC, 50 / 60 Hz. Auto select
Sample inlet connection ¼-inch NPT female, ¼-inch compression fitting tubing (included)
Sample outlet (drain) connection ⅜-inch NPT female, ⅜-inch hose barb tubing (included)
Sample inlet tubing ¼-inch OD or 6 mm OD
Sample outlet tubing ⅜-inch OD or 9 mm OD
Turbidimeter body drain Quick connect with integrated check valve
Analogue output:
Measurement module
1 selectable 0 – 20 mA or 4 – 20 mA; Output span programmable over any portion of the measurement range
Analogue output:
Junction box
1 selectable 0 – 20 mA or 4 – 20 mA; Output span programmable over any portion of the measurement range
Alarms
(Requires Junction Box option)
Three set-point alarms, each equipped with an SPDT relay with unpowered contacts rated 5A resistive load at 230 VAC
Digital Protocol options
(Requires Junction Box option)
Modbus, Profibus or Ethernet
Enclosure type:
Junction box
Fiber reinforced polyester
Enclosure rating Junction box: IP66
Measurement module: IP65
Compliance PTV 1000 IR (ISO): ISO 7027
PTV 1000 WL (EPA): Pending
PTV 2000 RL (EPA): Pending
Safety Pending
Immunity Pending
Emissions Pending
Mounting Hardware Turbidimeter Sensor – slotted mounting bracket that can be affixed to any vertical surface or panel (optional).
Junction box – direct mounting to any vertical surface or panel (optional)
Dimensions
PTV sensor with junction box
334.5 x 158.5 x 340.4 mm (L x W x H)
Shipping weight < 5 kg
Method of calibration One point calibration at 5.0 or 20 NTU with any regulatory approved formazin
Method of verification Wet standards or dry verification device

Delivery Content:

  • Turbidimeter
  • Mounting hardware for panel or wall mount
  • T-CALplus Kit – specify range – 1.0, 0.3, 20.0 or 5.0 NTU
  • Operator’s manual

 

PTV_Series_Literature_June 17